Model
LCR-2100
50Hz-100kHz, 34 points
LCR-2200
20Hz-200kHz, 15025 points,
Basic measurement accuracy (See details in technical specification)
LCRZ
0.05%
DCR, turn ratio
0.1%
Calibration condition
Warm up time:≥ 30 s
Environment temperature:23±5ºC
Signal voltage:0.3Vrms – 1Vrms
Zero clearing:after OPEN, SHORT
Testing cable length:0 m
Test signal frequency
LCR-2100
50Hz-100kHz , 34points
LCR-2200
20Hz-200kHz, 15025 points
Signal source output impedance
Selectable 30Ω, 100Ω, ±1% @1kHz
AC test signal level
Normal
10mV—2Vrms
(LCR-2100:
50mV—1Vrms)
Minimum resolution:10mV,
Accuracy:10% x setting voltage+2mV
100μA—20mArms
(TH2830:
500μA—10mA rms)
Minimum resolution:0.1mA
Constant level
20mV—1Vrms
(LCR-2100
100mV—500mVrms)
Minimum resolution:10mV
Accuracy:10% x setting voltage +2mV
200μA—10mArms
(LCR-2100
1mA—5mA rms)
Minimum resolution:0.1mA
DC bias voltage source
0V— ± 5V
Minimum resolution:0.5mV, Accuracy:1% x setting voltage +5mV
0mA—± 50mA
Minimum resolution:0.5μA
Transformer test parameters
Turn Ratio, Turns, Phase, L, C, Lk, Q, ACR, DCR, Balance, Pin Short
LCR test parameters
|Z|, |Y|, C, L, X, B, R, G, D, Q, θ, DCR
DCR display range
0.00001 Ω-999.999 MΩ
DCR polarity configuration
FIX, ALT
Turn ratio display range
1:0.001-1000:1
LCR parameters display range
|Z|, R, X 0.00001Ω — 99.9999MΩ
|Y|, G, B 0.00001μs — 99.9999s
C 0.00001pF — 9.99999F
L 0.00001μH — 99.9999kH
D 0.00001 — 9.99999
Q 0.00001 — 99999.9
θ(DEG) -179.999º — 179.999º
θ(RAD) -3.14159 — 3.14159
Δ% -999.999% — 999.999%
Measurement time (≥10 kHz)
Fast: 75 meas/sec(13ms), Medium:11 meas/sec(90 ms) Slow:2.7meas/sec(370 ms)
Equivalent circuit
Serial, Parallel
Range mode
Auto, Hold
Trigger mode
Internal, Manual, External, Bus
Average time
1-255
Zero clearing
Open, Short, Load
Arithmetical operation
Direct reading, ΔABS, Δ%
Delay time setting
0 -- 60.000s, 1ms steps
Comparator function
10 bins, BIN1~BIN9, NG, AUX
Bin count function
PASS, FAIL front panel or scan fixture LED display
Memory
100 groups of LCRZ instrument setting files
201 test result
Interface
Control interface
HANDLER or SCANNER
Communication interface
USB HOST, RS232C, RS485(option), GPIB(option)
Storage interface
USB DEVICE (U-disk storage or barcode scan)
LCR-2100 100KHz LCR meter
LCR-2200 200KHZ LCR Meter
Model | LCR-2100 | LCR-2200 |
---|---|---|
Basic accuracy
|
0.05% | 0.05% |
Test speed
|
13 Time/Second | 13 Time/Second |
Display screen
|
4.3 Inch TFT LCD Display | 4.3 Inch TFT LCD Display |
Test Voltage
|
50mV- 2V | 50mV- 2V |
Test Frequency
|
50Hz - 100KHz | 20KHz -200KHz |
Signal impedance
|
30/100 Ohm | 30/100 Ohm |